Compensation filters allow increased visibility of detail in chest, shoulder, spine, hip, knee, and foot radiographs. This study examines use of an anatomic compensation filter to improve imaging in pedal radiographs. Anteroposterior radiographs were obtained of 25 cadaveric feet at two settings with and without the compensation filter. Densitometer readings were taken at ten forefoot anatomic sites. The compensation filter produced statistically significant reductions in densitometer readings at all anatomic sites and at both radiographic settings. Filtration improved imagery of bony structures, provided excellent soft-tissue visualization, and lowered patient exposure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Podiatric Medical Association
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.