Abstract

Purpose:This work examines the dosimetric performance of two algorithms creating a corrected CBCT (corrCBCT) and a virtual CT (vCT) implemented in a commercial treatment planning system. Methods:60 patients distributed across all patient groups treated with curative intent at Vejle Hospital (breast, lung, prostate and anal/rectal cancer) were selected for the present study. Clinical treatment plans were recalculated on corrCBCT and vCT, as well as a reference CT (refCT) acquired as close in time to the CBCT image as possible. Recalculated doses were compared using gamma analysis, as well as by comparing D98%, D50%, and D2% for all delineated targets and organs at risk. Results:High dosimetric accuracy is demonstrated on both the corrCBCT and vCT. Gamma 2%/2mm pass rates >98% were found for all patients except two outliers still having >93% pass rates. Equivalence of all evaluated dose metrics within ±1Gy was observed for all patient groups, while the pelvic patients additionally showed equivalence for all metrics within ±1% of the refCT dose. For the thoracic patients, equivalence within ±2.5% was established for all metrics except median dose to the ipsilateral lung, calculated on corrCBCT for the breast patient group. Conclusion:The corrCBCT and vCT images are shown in excellent dosimetric agreement with refCT images, and show high potential for future use for streamlined adaptive radiotherapy workflows.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.