Abstract

Nowadays, a rotary blood pump can be used as not only for a bridge to transplantation (BTT) but also for a bridge to recovery (BTR) and a destination therapy (DT). In such cases, evaluation of the recovery level of the native heart provides useful information to improve the clinical strategy and decide adequate timing for removing of the RBP. In contrast, the indices for cardiac function have been studied. However, most of them do not consider the assistance with the RBP. In this study, we aimed at evaluating whether Emax, which is an index for cardiac function based on the pressure-volume relationships, is still valid during assistance with the RBP from an animal experiment. In the acute animal experiment with an adult goat, we measured pressure-volume (P-V) loops while cardiac function was normal, augmented or diminished. The experimental results revealed that there were typical differences in the shapes of P-V loops when the cardiac function was altered, and Emax can still be used as an index for the cardiac function even if the assistance with the RBP is ongoing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.