Abstract

Several models in the literature correlate blood pressure (BP) with the electrocardiogram (ECG) and photoplethysmogram (PPG) signals. These studies show substantial differences and have not been adequately compared. The MIMIC database, containing intrusive BP measures collected in a hospital setting, has been used to perform an extensive study of different models and variables extracted from the PPG and ECG signals.The best BP estimator model obtained is α/PAT2+βHR+δ, where PAT is the Pulse Arrival Time and HR is the heart rate. This model allows the monitoring of fast BP trends and fluctuations. If absolute values of BP were required, the model would have to be calibrated with real BP measures. The model, with periodic recalibrations, meets the Association for the Advancement of Medical Instrumentation (AAMI) requirements for diastolic BP but not for systolic BP, for which the mean error is close to 8 mmHg.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.