Abstract

BackgroundMatrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) is now widely used to detect pathogens in clinical settings in Japan.MethodHere, we report the effects of adding lysis buffer in the MALDI-TOF MS method to directly detect bacteria from 3 blood culture systems and compare their detection efficiencies for each pathogen.ResultsBlood culture broths from BD, bioMérieux, and Oxoid showed similar detection efficiencies without lysis buffer use and Escherichia coli, Klebsiella pneumoniae, and Pseudomonas aeruginosa were efficiently detected in all broths when lysis buffer was used. However, Streptococcus pneumoniae was not detected in BD broth when lysis buffer was added. Furthermore, Haemophilus influenzae and Bacteroides fragilis were not detected in all 3 systems when lysis buffer was used.ConclusionOptimization of blood culture system and lysis buffer is necessary according to each pathogen for direct identification by MALDI-TOF MS methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.