Abstract
A feasibility study was undertaken to evaluate laboratory phlebotomists performing bedside glucose monitoring (BGM) over a 3-month period on a medical and surgical floor. Specific questions included: feasibility of providing testing on a 24-h basis, accuracy, appropriate utilization, effect on patient care, and an analysis of cost. In all, 1975 tests were performed on 114 patients. BGM results were within 15% of the laboratory's result 97% of the time. Patient and physician satisfaction was high. Although the cost of BGM is slightly higher than a laboratory glucose test, its use appeared to reduce the length of hospital stay by 0.47 days. Practical information on initiating a highly successful BGM program is provided.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.