The objective of this work is to develop an easy-to-build and robust setup for measuring the nonlinearity parameter B∕A in fluids using ultrasound. The method is based on the pulse-echo technique, using a single element broadband acoustic transducer, and requires electrical signal measurements. Results obtained in water and denatured alcohol validate the proposed procedure. The choice of a suitable primary wave frequency is discussed with regard to the transducer sensitivity. Further, the influence of the perturbations introduced by the experimental device nonlinearities, and the role of the reflector on the measured second harmonic field amplitude are investigated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.