Abstract

This study represents the first national survey conducted in Sri Lanka to establish national diagnostic reference levels (NDRLs) for screening and diagnostic acquisitions in digital mammography (2D-DM) and digital breast tomosynthesis (DBT). Additionally, the study investigated the relationship between average glandular dose (AGD) and compressed breast thickness (CBT) and introduced a novel concept called dose range bands (DRBs) as a tool for optimizing radiation dose in mammography. DICOM metadata was used to extract dose data and exposure parameters from women undergoing either screening (350) or diagnostic (750) DM. The analysis included both craniocaudal and mediolateral oblique views of each breast, acquired using 2D-DM and DBT imaging techniques. The NDRL (AGD per view) was 1.97 mGy and 2.01 mGy for diagnostic DM and DBT, respectively. The corresponding NDRLs for screening acquisition were 2.44 mGy and 2.30 mGy. The mean DBT/DM AGD ratio was 1.39 and 0.97 for diagnostic and screening, respectively. Further, the association between the average glandular AGD per view and CBT is stronger in DBT than in 2D-DM. The study findings highlight the need for standardisation of compression practices, considering factors such as the radiographer’s experience, imaging equipment, breast density, age, breast size, and pain threshold.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.