Abstract
The use of an inductively coupled plasma, ICP, as an excitation source for atomic fluorescence spectrometry, AFS, in a second ICP is re-examined. Improvements in the ICP-ICP-AFS setup have allowed the lowering of the limits of detection by one to two orders of magnitude below that of previous work. Also discussed is a new mode of operation for the atomization cell ICP. Through simple torch-position and flow-rate adjustments, a thin plasma, which extends 20 to 30 cm above the torch, can be produced. This plasma is referred to as the pencil plasma. With the use of these operating conditions, propane can be added to the Ar nebulizing gas to aid in refractory-element determination. The pencil plasma and the conventional plasma will be compared for use as an atom reservoir for AFS measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.