Abstract

A simple and rapid flow injection-arsenic hydride generation-atomic fluorescence spectrometric (FI-HG-AFS) method has been optimized for the direct determination of arsenic in untreated samples of beer and wine. Several aspects were evaluated: concentration of HCl and NaBH 4 , argon and hydrogen flow-rates and effect of oxidation state on the analytical signal. The optimum conditions selected were 6 m HCl and 0.5% (m/v) NaBH 4 and 6 m HCl and 1% (m/v) NaBH 4 to determine total arsenic contents of different brands of beer and wine, respectively. The accuracy of the method was evaluated by comparing the results obtained with those provided for three mineralization procedures: HNO 3 -H 2 O 2 , H 2 SO 4 -HNO 3 -HClO 4 and Mg(NO 3 ) 2 -MgO mixtures. The results obtained showed good agreement between the conventional digestion procedures and the direct determination of arsenic in the untreated samples. The method enables the determination of arsenic to be made in untreated samples with detection limits of 0.3 and 0.5 µg L –1 for beer and wine, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.