Abstract

Abstract Medical implants play a central role in modern medicine and both, naturally derived and synthetic materials have been explored as biomaterials for such devices. However, when implanted into living tissue, most materials initiate a host response. In addition, implants often cause bacterial infections leading to complications. Polyelectrolyte multilayer (PEM) coatings can be used for functionalization of medical implants improving the implant integration and reducing foreign body reactions. Some PEMs are also known to show antibacterial properties. We developed a PEM coating suggesting that it can decrease the risk of bacterial infections occurring after implantation while being highly biocompatible. We applied two different standard tests for evaluating the PEM’s antibacterial properties, the ISO norm (ISO 22196) and one ASTM norm (ASTM E2180) test. We found a reduction of bacterial growth on the PEM but to a different degree depending on the testing method. This result demonstrates the need for defining proper method to evaluate antibacterial properties of surface coatings.