Abstract

The iodide ion selective electrode has potential use for continuous monitoring of total oxidants in ambient air. Total oxidants are measured coulometrically or colorimetrically after oxidation of neutral buffered Kl solution (0.6-1.2M) to free iodine. On the other hand, the iodide electrode can measure much more dilute solutions (<10-4M). The iodide electrode measures the disappearance of iodide from the absorbing solution. The detector is made from an iodide electrode, reference electrode, an expanded scale pH meter, and millivolt recorder. The iodide electrode shows a constant Nernstian —59 mv change per decade over a dilute solution range although the absolute voltage shows a drift of 1 mv/day. A continuous iodide electrode air monitor was set up and investigated for ambient monitoring. Operational parameters as air and solution flow rates and electrode parameters as response time at different iodide concentrations were investigated. Statistical comparison was made with a Beekman colorimetric total oxi...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.