Abstract
A gradient LC method for the determination of related substances in ritonavir (RTV) has been recently published in the International Pharmacopoeia. The method uses a base-deactivated reversed-phase C18 column kept at a temperature of 35 °C. The mobile phases consist of acetonitrile, phosphate buffer pH 4.0 and water. UV detection is performed at 240 nm. A system suitability test (SST) is described to govern the quality of the separation. Since no brand names of columns are mentioned in pharmacopoeial texts, analysts often have problems to select a suitable stationary phase which is only described in general terms. So, the separation towards RTV components was investigated on 18 C18 columns and correlation was made with the column classification system developed in our laboratory. The method was further evaluated using a Hypersil BDS C18 column (25 cm × 4.6 mm i.d.), 5 μm, which was also used for the development of the method. A central composite design was applied to examine the robustness of the method. The method shows good precision, linearity, sensitivity and robustness. Four commercial samples were examined using this method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.