Abstract

A new cubic phantom was designed to improve the efficiency on the QA measurement of accelerator. It has a variety of applications, such as dose constancy check, depth dose verifications, and symmetry and flatness evaluations. In particular, this new design makes it much easier to perform the check on output constancy vs. gantry angles as a cylindrical ion chamber positioned at the center of the phantom. The reproducibility of the setup using this phantom has been investigated. The charge effect of the phantom is found to be insignificant. It also reduces the monthly dosimetric QA time spent on a typical multimodality accelerator by approximately 40%. PACS number(s): 87.66.–a

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.