BACKGROUNDCulture-based systems are currently the preferred means for bacterial screening of platelet (PLT) concentrates. Alternative bacterial detection techniques based on nucleic acid amplification have also been developed but these have yet to be fully evaluated. In this study we evaluate a novel 16S rDNA polymerase chain reaction (PCR) assay and compare its performance with automated culture.STUDY DESIGN AND METHODSA total of 2050 time-expired, 176 fresh, and 400 initial-reactive PLT packs were tested by real-time PCR using broadly reactive 16S primers and a “universal” probe (TaqMan, Invitrogen). PLTs were also tested using a microbial detection system (BacT/ALERT, bioMérieux) under aerobic and anaerobic conditions.RESULTSSeven of 2050 (0.34%) time-expired PLTs were found repeat reactive by PCR on the initial nucleic acid extract but none of these was confirmed positive on testing frozen second aliquots. BacT/ALERT testing also failed to confirm any time-expired PLTs positive on repeat testing, although 0.24% were reactive on the first test. Three of the 400 “initial-reactive” PLT packs were found by both PCR and BacT/ALERT to be contaminated (Escherichia coli, Listeria monocytogenes, and Streptococcus vestibularis identified) and 14 additional packs were confirmed positive by BacT/ALERT only. In 13 of these cases the contaminating organisms were identified as anaerobic skin or oral commensals and the remaining pack was contaminated with Streptococcus pneumoniae.CONCLUSIONThese results demonstrate that the 16S PCR assay is less sensitive than BacT/ALERT and inappropriate for early testing of concentrates. However, rapid PCR assays such as this may be suitable for a strategy of late or prerelease testing.
Highlights
Culture-based systems are currently the preferred means for bacterial screening of platelet (PLT) concentrates
We evaluate the performance of this ethidium monoazide (EMA)-enhanced real-time 16S rDNA polymerase chain reaction (PCR) for screening of PLT concentrates in parallel with BacT/ALERT automated culture
The sensitivity of the EMA-enhanced 16S rDNA real-time PCR was confirmed in each assay run with a 10-fold dilution series of E. coli DNA representing 5000, 500, 50, and five genome copies per reaction
Summary
Culture-based systems are currently the preferred means for bacterial screening of platelet (PLT) concentrates. PLTs were tested using a microbial detection system (BacT/ALERT, bioMérieux) under aerobic and anaerobic conditions. RESULTS: Seven of 2050 (0.34%) time-expired PLTs were found repeat reactive by PCR on the initial nucleic acid extract but none of these was confirmed positive on testing frozen second aliquots. BacT/ALERT testing failed to confirm any time-expired PLTs positive on repeat testing, 0.24% were reactive on the first test. Three of the 400 “initial-reactive” PLT packs were found by both PCR and BacT/ALERT to be contaminated (Escherichia coli, Listeria monocytogenes, and Streptococcus vestibularis identified) and 14 additional packs were confirmed positive by BacT/ALERT only. CONCLUSION: These results demonstrate that the 16S PCR assay is less sensitive than BacT/ALERT and inappropriate for early testing of concentrates. Rapid PCR assays such as this may be suitable for a strategy of late or prerelease testing
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.