An emergency cricothyrotomy device was placed in 11 anesthetized dogs in order to assess airway damage and problems in placement, ventilation, and design. Posterior airway perforation without esophageal damage occurred in three animals. Submucosal hematomas or cricoid cartilage injury occurred in seven animals. As placement by an untrained operator was surprisingly difficult, prior operator training with this device is recommended. When properly installed, an excellent emergency airway was achieved. Some minor design shortcomings are noted and suggestions for improvement are made.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.