Abstract

The objectives of this study were to evaluate an automated device for ventilatory support based on AMBU manufactured in March 2020. The ESSI-1 INC was evaluated through pulmonary mechanics and physiology parameters through compensatory spirometer tests (TISSOTs), and an artificial lung Model5600i Dual Adult PNEU VIEW SYSTEM; it was also compared to the anesthetic ventilatory support equipment (AEONMED 7500) in porcine models, measuring ventilatory, hemodynamic and gasometric parameters. This equipment (ESSI-1 INC) was successfully tested by mechanical and biological models, such as pigs in which its performance was evaluated in terms of variability of tidal volume, ventilation frequency, and I/E relationship versus the manual performance of two medical interns. All the results turned out as expected and were satisfactory. It is safe and effective equipment and should be tested and used in diverse clinical conditions to standardize the ventilatory safety and care of patients who require it.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.