The findings from a preliminary assessment of a new instrument designed for the inoculation and spreading of specimens for microbiological analysis onto agar plates are described. The study found that the instrument was able to select full or biplates from a number of input cassettes, each containing different agar types. Samples were then inoculated by the instrument onto the agar surfaces and spread by a novel plastic applicator. Following this, the instrument labeled the plates and sorted them into a number of specified output stations. It was found that the instrument was able to inoculate and spread samples over a greater proportion of the agar plate surface than the manual loop-to-plate method. As a consequence, up to 44% more usable colonies were produced per plate from clinical specimens and standard cultures. Viable counts showed that the instrument was able to detect as few as 10(2) CFU/ml in fluids and also facilitated the enumeration of organisms, particularly in specimens such as urine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.