In present work, an alternative irradiation system based on a symmetric cylindrical tank filled with a moderator containing hydrogen, which was equipped with a NaI(Tl) scintillation detector, was proposed for using in determination of neutron flux. This irradiation system was designed by MCNP4C code, with considering a 241Am–Be neutron source in several volumes and different materials. When the neutron is captured by hydrogen, a 2.22 MeV prompt gamma-ray is emitted. The gamma pulse-height spectrum shows a photo-peak around 2.22 MeV whose net area is proportional to the total emission rate of neutron. The simulation result showed that a cylindrical tank with 110 cm diameter and height filled with water can be a suitable system for neutron source strength calibration. Furthermore, a proper two-layer shielding must be placed between the source and detector for preventing neutrons and gamma rays to directly enter the detector.