Abstract

AbstractDelamination of thin film polymeric coatings from metallization layers is a common cause of failure in biomedical implants. To address the problem, different adhesion promotion techniques can be applied which include surface pre-treatment with oxygen and argon plasma and the use of different adhesion promoters. In this paper the applicability of titanium (Ti), silicon oxide (SiOx), diamond-like carbon (DLC), tetramethylsilane (TMS) and aluminium oxide (AlOx) as adhesion promoters is evaluated. A cross cut, peel and scratch test are used to qualify and quantify the adhesion before and after storage in phosphate buffered saline (PBS) for 48 hours at a temperature of 37 °C. Promising results could be achieved by a combination of Ti and DLC as well as by AlOx.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.