Abstract

The stability of two commonly used uncemented acetabular prostheses fixed with either pegs or screws has been established by evaluating the torque strengths in ten cadaveric acetabulae. The initial stability of implants to simulated frictional loads was satisfactory compared to previous estimations of the frictional torques in polyethylene-on-metal and metal-on-metal articulations. Testing of the torque required to produce 2 o of rotation at the bone-implant interface showed that the peg fixation required significantly greater loads than did the screw fixation ( P<.03). Testing the prostheses to failure gave good results, although lower than previous cemented arthroplasty estimations. The margin of stability over the frictional torques that can be generated is such that it may affect the bony ingrowth into these prostheses. This may account for the common finding of fibrous ingrowth in these prostheses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.