Abstract

A 2021 publication in Biosensors from Moser and colleagues [...].

Highlights

  • A 2021 publication in Biosensors from Moser and colleagues [1] reported on the accuracy of the FreeStyle Libre® intermittently scanned continuous glucose monitoring system at different rates of change in plasma glucose (RCPG), during two oral glucose tolerance tests (OGTT) in 19 adults with type 1 diabetes (T1D), following 12 and 36 h fasting periods

  • They concluded that the FreeStyle Libre system is accurate when compared to reference plasma glucose during OGTT glycemic challenge, but that its performance was dependent on the RCPG and that it overestimated glucose during periods when reference blood glucose values were in hypoglycemia

  • The most recent algorithm for the FreeStyle Libre sensor does not report the mean absolute relative difference (MARD) for readings exclusively

Read more

Summary

Introduction

A 2021 publication in Biosensors from Moser and colleagues [1] reported on the accuracy of the FreeStyle Libre® intermittently scanned continuous glucose monitoring (isCGM) system at different rates of change in plasma glucose (RCPG), during two oral glucose tolerance tests (OGTT) in 19 adults with type 1 diabetes (T1D), following 12 and 36 h fasting periods. They concluded that the FreeStyle Libre system is accurate when compared to reference plasma glucose during OGTT glycemic challenge, but that its performance was dependent on the RCPG and that it overestimated glucose during periods when reference blood glucose values were in hypoglycemia.

Results
Conclusion