Computer programs for calculation of the backscatter factor and absorbed dose in breast phantoms are developed based on Monte Carlo simulation. Rad/roentgen conversion factors are calculated for water, fat, a mixture of 50% water and 50% fat, and lucite irradiated with monoenergetic and polyenergetic x rays from a tungsten or molybdenum anode x-ray tube. These factors can be used to estimate skin dose, average depth dose, or average integral dose in a breast when the beam quality, exposure in air at the skin position, and composition of the breast are known. Calculated backscatter factors are considerably greater than measured values reported previously.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.