Abstract
PurposeTo evaluate a two‐dimensional diode array for patient‐specific quality assurance of VMAT stereotactic radiosurgery (SRS) plans.MethodsThe diode array, an SRS MapCHECK (SRSMC), was composed of a 77 mm ×77 mm face‐centered array having a spacing of 2.47 mm. Sixty SRS plans were selected from our clinical database, 30 for treatment of a single target and 30 for multiple targets. The target sizes ranged from 2.4 mm to 44.7 mm equivalent diameter (median 8.7 mm). The plans were delivered to the diode array. For multiple target plans, two measurements were obtained at two locations, one corresponding to the largest target and the other to the smallest target. Gamma using a 3%/1 mm criteria and the dose to the center diode were compared with radiochromic film (RCF). Dose to selected regions of the detector electronics was calculated.ResultsThe mean difference between the center diode and RCF was −1.2%. For a threshold of at least 95% of detectors/pixels having gamma < 1 for a 3%/1 mm criteria, SRSMC and RCF gave consistent results for 79 of the 90 measurements. For plans with an arc having a patient support angle of 90° or 270°, the median dose to the electronics was 0.65% of the prescription dose.ConclusionsSRSMC is an efficient tool for accurate patient‐specific quality assurance of VMAT single and multiple target radiosurgery, yielding similar clinical decisions as radiochromic film.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.