Abstract

A Dynamic Reaction Cell™ inductively coupled argon plasma mass spectrometer (DRC-ICP-MS) was evaluated for the determination of chromium and vanadium in serum and urine. Reaction cell conditions were evaluated for the elimination of ArC + and ClOH + interferences on chromium at mass 52 and OCl + on vanadium at mass 51. A diluent containing only 1% nitric acid and internal standards (Y and Ga) was used to prepare serum and urine for analysis. Instrument response calibration was achieved by using aqueous acidic standards spiked into pooled sera or urine matrices. The slopes of the calibration curves prepared in urine and serum matrices were nearly identical. On average, chromium detection limits are 2.5 times lower using the DRC than Zeeman graphite furnace atomic absorption spectrometry (ZGFAAS). Vanadium detection limits are approximately 50 times lower. Average detection limits achieved with DRC-ICP-MS are 0.075 μg Cr/l and 0.028 μg V/l. Average results for the analysis of National Institute of Standards and Technology Standard Reference Material (NIST SRM) 1598 Bovine Serum (attained over 22 days) are: 0.14 μg Cr/l and 0.068 μg V/l. The reference concentrations for vanadium and chromium in NIST SRM 1598 are (0.06) μg V/l and 0.14±0.08 μg Cr/l, respectively. Results for chromium and vanadium determinations on ICP-MS survey samples from the Toxocologie du Quebec are equivalent to those reported by high resolution inductively coupled plasma mass spectrometry (HR-ICP-MS) for the same survey samples.