Abstract
This is a study of the reaction of large nerves to implantation using a flexible, thin-film cuff electrode. Cuff electrodes were implanted on the sciatic nerve of three cats. An implantation period of six weeks allowed sufficient time for any injury responses in the nerve and connective tissue sheath around the cuff to develop. The electrode came off the nerve in one of the cats. In the remaining two cats, gross observation following explantation of the electrodes revealed encapsulation of the cuffs without swelling of nerve tissue. Histological evaluation did not demonstrate nerve injury. The nerve cuff electrodes, which are comprised of titanium and iridium coatings on a fluorocarbon polymer substrate, appeared unaffected by the implantation, and connective tissue encapsulation did not adhere to either the polymer substrate or metallization. Evaluation of the electrodes using activated iridium oxide charge injection sites in more extended studies is now being undertaken.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.