Abstract

A second-generation ultra-fast energy-resolved application specific integrated circuit (ASIC) has been developed for photon-counting spectral computed tomography (CT). The energy resolution, threshold dispersion and gain of the ASIC were characterized with synchrotron radiation at Diamond Light Source. The standard deviation of threshold offsets at zero keV is 0.89 keV. An RMS energy resolution of 1.09 keV has been demonstrated for 15 keV photon energy at a count rate of 40 kcps, and it deteriorates at a rate of 0.29 keV/Mcps with the increase of output cout rate. The count rate performance of the ASIC has also been evaluated with 120 kV polychromatic x-rays produced by a tungsten anode tube and the results are presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.