Abstract

The purpose of this study was to investigate the potential of real-time optically stimulated luminescence (rtOSL) measurements of a beryllium oxide (BeO) ceramic fibre-coupled luminescence dosimetry system. By pulsing the stimulation laser during the exposure to ionizing radiation, an rtOSL dose-rate measurement can be obtained which could be stem effect free. A portable rtOSL BeO ceramic fibre-coupled dosimetry system is presented and characterized using a constant dose-rate superficial 140 kVp X-ray beam. The rtOSL was measured for dose-rates between 0.29 and 3.88 Gy/min, controlled by varying the source to surface distance. After correcting for OSL decay during the exposure, a linear dose-rate response of the change in rtOSL (ΔrtOSL) was observed. The ΔrtOSL was also observed to be stem effect free.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.