Abstract

To evaluate the accuracy of a real-time, continuous glucose monitoring system (CGMS) in healthy dogs undergoing anesthesia for elective ovariohysterectomy or orchiectomy. 10 healthy dogs undergoing routine elective surgery. A CGMS was placed and used to obtain calculated glucose measurements before, during, and after anesthesia in each dog. Periodically, CGMS measurements were compared with concurrent measurements of glucose concentration in peripheral venous blood obtained with a portable chemistry analyzer (PCA). CGMS-calculated glucose measurements were significantly different from PCA blood glucose measurements during most of the anesthetic period. The CGMS values differed from PCA values by > 20% in 54 of 126 (42.9%) paired measurements obtained during the anesthetic period. Hypoglycemia was evident in CGMS measurements 25 of 126 (19.8%) times during anesthesia. By comparison, only 1 incident of hypoglycemia was detected with the PCA during the same period. Use of the CGMS for routine monitoring of interstitial glucose concentration as an indicator of blood glucose concentration during anesthesia cannot be recommended. Additional investigation is necessary to elucidate the cause of discrepancy between CGMS results and PCA data during anesthesia.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.