Abstract

A rapid latex agglutination test (Bactigen Group B Streptococcus Cervical Screen) for detection of group B streptococci in cervical-vaginal specimens was evaluated using two different slide systems, the traditional serologic slide and capillary action track (Trak) slide. Culture was used as reference method. A total of 344 cervical-vaginal specimens were tested. The group B streptococci carrier rate was found by culture to be 10.8%, 56.8% of these specimens being heavily colonized. The sensitivity and specificity of the latex agglutination test in heavily colonized specimens was 95.2% and 99.3% for the serologic and track slides respectively. The overall sensitivity, including lightly colonized specimens, was 62.2%. The positive predictive value was 92% for both slide systems, and the negative predictive value 95.4% and 95.6% for the serologic and track slides respectively. The latex agglutination test, used with either slide, provides a rapid and effective method for identification of specimens heavily colonized with group B streptococci. The track slide may provide a convenient alternative to serologic slides since it does not require rotation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.