Abstract

To evaluate a point-of-care anticoagulant rodenticide lateral flow analyzer for the detection of various rodenticide compounds. Prospective, laboratory study. University teaching hospital. The study utilized a serum sample from one healthy canine donor. Samples were centrifuged and serum samples were aliquoted and either used within 4 hours or frozen at -70°C for further quantitative analysis. Samples were spiked with clinically relevant concentrations of 1 of 6 rodenticide compounds (warfarin, pindone, chlorphacinone, brodifacoum, bromethalin, and its metabolite desmethylbromethalin). Seventy-five microliters of spiked serum (or unaltered serum) was introduced into the lateral flow test. Three readers who were blinded to the sample preparation interpreted the lateral flow test as either positive or negative for the presence of anticoagulant rodenticide. All readers were in agreement for the results of each serum sample. The point-of-care test kit was able to detect a single anticoagulant rodenticide (warfarin) at concentrations below the manufacturer's recommended limit of detection, but was unable to detect any other anticoagulant rodenticide. The results of this test and therapeutic interventions must be considered in light of history, physical examination, and other clinical data. Based on results from this study, the test kit only detects warfarin and not other more common second-generation anticoagulant rodenticides.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.