Abstract

Venous thromboembolism (VTE) is the fourth most common preventable hospital-acquired complication for hospitalized trauma patients. Mechanical prophylaxis, using sequential compression or intermittent pneumatic compression (IPC) devices, is recommended alongside pharmacologic prophylaxis for VTE prevention. However, compliance with device use is a barrier that reduces the effectiveness of mechanical prophylaxis. This study aimed to determine whether using the Movement and Compressions (MAC) system compared with an IPC device impacts compliance with mechanical VTE prophylaxis in trauma patients. This study used a before-and-after design with historical control at a Level II trauma center with a convenience sample of adult trauma patients admitted to the intensive care unit or acute care floor for at least 24h. We trialed the MAC device for 2 weeks in November and December 2022 with prospective data collection. Data collection for the historical control group occurred retrospectively using patients from a point-in-time audit of IPC device compliance from August and September of 2022. A total of 51 patients met inclusion criteria, with 34 patients in the IPC group and 17 patients in the MAC group. The mean (SD) prophylaxis time was 17.2h per day (4.0) in the MAC group and 7.5h per day (8.8) in the IPC group, which was statistically significant (p < .001). Our findings suggest that the MAC device can improve compliance with mechanical prophylaxis.