Abstract
The Los Alamos National Laboratory tritium processing facility purchased an ANTECH Series P300 Isothermal Tritium calorimeter from the ANTECH Corporation in 2005. The instrument is used to nondestructively measure the quantity of tritium in an item based on the heat output generated by tritium beta decay. Instrument performance data collected over the past two years is presented showing that this instrument can detect as low as 0.001 W (0.003 g of tritium) within rather large measurement canisters of approximately 170 mm (6.8 in.) diameter by 610 mm (24 in.) long. With a manufacturerstated measurement power range of 0.005 W to 15 W, this calorimeter has performed beyond the specified purchase requirements. Using a combination of sensors for its thermal element, the P300 calorimeter has demonstrated exceptional sensitivity and precision.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.