AbstractAn improved wall‐jet configuration type flow‐through cell has been developed for use with commercial ion‐selective electrodes (ISEs) and has been evaluated in the flow injection potentiometric (FIP) mode. The flow‐through cell was designed to accommodate commercial ISEs that employ a range of sensing surfaces, including gas‐sensing, liquid polymer membrane and solid‐state based ISEs. Commercial ISEs employed for the evaluation of the flow‐through cell were ammonia, calcium and iodide ion‐selective electrodes. The observed electrode slopes were −58.8 ± 0.9, 25.2 ± 0.5 and −58.3 ± 0.9, mV change / activity decade, respectively, over a wide log‐linear range using a 100 μL sample volume in the FIP mode. Sample throughput rates as fast as 240 injections/h and analytically useful peak height responses can be achieved using injected sample volumes as low as 25 μL with all three types of ISEs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.