Abstract

A new reentrant ionization chamber SNC 1008 has been introduced for low dose rate (LDR) and high dose rate (HDR) brachytherapy source calibrations. There is no air density correction requirement for the chamber since it is hermetically sealed. In this study we evaluate its use for HDR as well as LDR source calibrations. The magnitude of the ionization current collected was found to be independent of the sign of the polarizing voltage within +/- 0.5%. The axial response curve of the well chamber reveals a "plateau region" (+/- 0.5% dose variation) of 4.2 cm, which is larger than the published values for other commercially available chambers. Long-term stability of the chamber was evaluated. Short-term response and source positioning reproducibility were tested using both LDR and HDR sources. Ion collection efficiency was found to be high, making the chamber suitable for HDR 192Ir sources.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.