Abstract

Preparative HPLC-MS is often the method of choice for purification of small amounts (<100mg) of diverse new molecules, such as compound libraries for drug discovery. The method is robust, well proven, and widely applicable. In contrast, preparative supercritical fluid chromatography coupled with mass spectrometry (SFC-MS) has seen only slow acceptance for the same application--despite some potential scientific and economic advantages. One of the reasons for slow adoption of SFC-MS is the lack of well-proven, robust, and commercially available instrumentation. In early 2009, TharSFC (a Waters Company, Pittsburgh, PA) introduced a new fully integrated system for preparative SFC-MS: The SFC-MS Prep-100. We report herein an objective evaluation of the SFC-MS Prep-100, including tests for pump and autosampler performance, sample recovery, sample carryover, fraction triggering, detector/fraction collector synchronization, and overall robustness. Our results suggest that the SFC-MS Prep-100 represents a significant advance over previous generation instrumentation.