Abstract

A new therapeutic plasma exchange device developed by Sarns Inc./3M was evaluated in plasmapheresis of 20 healthy volunteers and in a multicenter clinical study of therapeutic plasma exchange that included 49 patients. Safety and efficacy of plasma separation from whole blood were assessed for a module that contains Durapore microporous surfactant-free polyvinylidene fluoride membrane (Millipore Corp., Bedford, Mass., USA). The extra-corporeal volume was 80 ml. Citrate and heparin anticoagulants were utilized. Mean plasma separation efficiency was 62% with unhindered passage of plasma proteins through the membrane pores and no hemolysis or activation of complement as measured by total hemolytic complement (CH50) and C3 conversion. Mean decrease in platelet count after procedures was 10%. No severe reactions occurred, and citrate effects (13%) were comparable to values reported with centrifugal instruments. The Sarns Inc./3M Therapore device is a rapid, safe and efficient system for plasma exchange and potentially for source plasma collection. The principal benefits are small extracorporeal volume and cell-free filtrate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.