Abstract
The LG HIV Ag–Ab Plus, a new fourth generation diagnostic assay for HIV infection, was evaluated in comparison to the Enzygnost HIV Integral, an established fourth generation HIV assay. The LG assay showed 100% sensitivity with 109 samples with anti-HIV-1, anti-HIV-2 or anti-HIV-1 group O reactivity. It also detected correctly all 51 positives on three BBI performance panels, slightly outperforming the Enzygnost HIV Integral, which detected 50. The specificity of the LG HIV Ag–Ab Plus was 99.9% with 999 sera from healthy blood donors, which was slightly inferior to the performance of the Enzygnost HIV Integral, which had 100% specificity. The LG assay showed 100% specificity with 81 specimens with underlying diseases including hepatitis B, demonstrating a low risk of cross-reactivity with other infections. The reduction of the diagnostic window by the LG HIV Ag–Ab Plus, compared to a third generation HIV assay, was 6.3 days. The LG assay also showed sufficiently high intra-person and inter-person reproducibility. The overall performance of this new fourth generation HIV assay was adequate for screening and diagnosis of HIV infection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.