Abstract

We have designed a new effect-site controlled, patient-maintained sedation system for delivering propofol. In the previous systems we developed, the patients retained the use of the handset throughout the procedure and were able to increase the level of sedation. However, it was found that this could potentially lead to oversedation. In the present system, the patients were able to increase their level of sedation until a level was reached that was judged by the patients as being adequate to allow them to tolerate the injection of dental local anaesthetic. The handset was then taken from the patients and the effect site concentration of propofol was maintained at that level for the remainder of the procedure. To assess its safety and efficacy, the system was used to sedate 40 patients presenting for dental procedures under sedation. The system was used successfully and treatment was completed in 39 patients. The system was found to be safe. Both surgeon and patient approval scores were high. Although this study demonstrates the efficacy of effect-site controlled, patient-maintained propofol sedation in this group of patients, further work is required to confirm its safety.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.