Abstract

ObjectivesA newly developed fully automated Lumipulse G AMH method (Fujirebio Diagnostics) was recently introduced in clinical laboratories for quantitative determination of anti-Müllerian hormone (AMH) level in human serum or plasma. AMH has emerged as value-added biomarker in the assessment of ovarian reserve, in diagnosis of granulosa cells cancer and in the investigation of gonadal disorders. We compared Lumipulse G AMH assay performances with other methods largely applied for AMH measurements. Design and MethodsThe Lumipulse G AMH method based on two-step sandwich chemiluminescence enzyme immunoassay was assessed on Lumipulse G600II analyzer. The evaluation study included imprecisions, sensitivity and linearity whereas a comparison study was performed on a heterogeneous population of 114 patients by using the Elecsys AMH Plus assay on COBAS 8000 e602 module (Roche Diagnostics). ResultsLumipulse G AMH system showed good repeatability (within-run imprecision) with CV values below 1% (0.5% and 0.9% for high and low serum pools). Similarly within-laboratory imprecision was assessed with CV values of 2.5% and 1.6% for high and low level controls respectively. A linearity regression formula of 1.0119x-0.067 with a coefficient of determination (r2) equal to 0.999 was obtained in a range from 0.044 to 22.42 ​ng/ml. Passing-Bablok regression analysis was performed for assay comparability of AMH measurements. Results were closely correlated (correlation coefficient ​= ​0.997) with a regression equation (y ​= ​1.230x-0.025) showing a positive slope. Also, Bland-Altman analysis confirmed a good agreement between Lumipulse G AMH and Roche Elecsys AMH Plus assays with a bias of 17.76% in a large measurement range. ConclusionsThe performance of Lumipulse G AMH system was highly comparable with that of Roche Elecsys AMH Plus assay although approximately 10% higher values of AMH levels were observed for Lumipulse AMH system at all range of concentrations. Nevertheless the Lumipulse G system seems to be largely suitable for quantitative determination of AMH level in small-scale laboratory because of the reduced size and the use of single cartridge per test assuring flexibility and easy handling.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call