Abstract
ObjectiveTo achieve higher tumor detection efficiency, we evaluated a multiplex assay for TM analysis based on the Luminex-100 multiplex suspension bead array. DesignThe assay simultaneously determined the concentrations of nine TMs in 1114 human serum specimens (546 patients with tumors, 158 patients with non-tumor inflammatory diseases, and 410 normal controls). The nine TMs were AFP, CEA, CA125, CYFRA 21‐1, CA242, f-PSA, t-PSA, NSE and free β-hCG. The multiplex suspension bead assays were compared with conventional methods used in clinical laboratories. ResultsThe Luminex assay has the same levels of sensitivity, specificity and accuracy in the prediction of positive tumor specimens as conventional methods. ConclusionMultiplex suspension bead arrays have promising applications in clinical laboratories.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.