Abstract

To evaluate the action of a metal artefact reduction (MAR) tool when artefact-generator metal object is at different positions in the field of view (FOV). A cylindrical utility wax phantom, with a metal alloy sample inside, was made. The phantom was positioned centrally and peripherally in the FOV for image acquisition, with and without the MAR tool activation. The standard deviation values (image noise levels) from areas around the metal sample and the control area were obtained. The numbers were compared by Student's t-test (α = 0.05). When the tool was activated, a significant difference of image noise was observed for central and peripheral positioning, for both control area (p = 0.0012) and metal area (p = 0.03), and a smaller level of noise was observed for images with phantoms in central positioning. A decrease in image noise with the tool activated was found only in phantoms with the metal object positioned centrally in the FOV. For the MAR tool to be effective, the artefact-generator object needs to be in the central region of the FOV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.