Abstract

Measuring sound-field auditory steady-state responses (ASSR) is a promising new objective clinical procedure for hearing aid fitting validation, particularly for infants who cannot respond to behavioral tests. In practice, room acoustics of non-anechoic test rooms can heavily influence the auditory stimulus used for eliciting the ASSR. To systematically investigate the effect of the room acoustics conditions on sound-field ASSR, a loudspeaker-based auralization system was implemented using a mixed order Ambisonics approach. The present study investigates the performance of the auralization system in terms of objective room acoustic measurements and sound-field ASSR measurements, both in the actual room and in the simulated and auralized room. The evaluation is conducted for a small room with well-defined acoustic properties. The room is carefully modeled using the novel room acoustic simulation tool PARISM (Phased Acoustical Radiosity and Image Source Method) and validated through measurements. This study discusses the limitations of the system and the potential improvements needed for a more realistic sound-field ASSR simulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.