Abstract

A gas jet-enhanced sputtering device has been utilized in glow discharge atomic absorption spectrometry for the analysis of solid steel samples. The experimental design was implemented in the evaluation of fundamental operating parameters (hollow cathode lamp current, burn current and burn pressure). Statistical techniques were used to determine the optimum conditions. Four samples types were considered: British Chemical Standard Certified Reference Materials SS 451/1–455/1 Carbon Steel Residual Series (Group A), SS 461–468 Austenitic Stainless Steels and SS 409/1 Low Alloy Steel (Group B), and 12671 H Cobalt-based Alloy. Sensitivities for nickel, chromium and copper determined in the carbon steel under optimum conditions were 54, 6 and 64 µg g–1, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.