Abstract

BackgroundHDV antibody testing is recommended for universal screening and as the first line in an HDV double reflex testing strategy for effectively identifying patients with active infection for therapeutic treatments. ObjectiveThe aim of this study is to evaluate the performance of a newly developed ARCHITECT HDV Total Ig (ARCHITECT HDV Ig) prototype assay. Study designPerformance characteristics were determined for the ARCHITECT HDV Ig and a reference test, LIAISON XL Anti-HDV using a well-characterized specimen panel, comprising HDV RNA positive (n = 62) and negative (n = 70) samples, and healthy US blood donors. ResultsHealthy US blood donors (n=200) showed 99.5% (199/200, 95%CI=97.65–99.98) specificity with ARCHITECT HDV Ig and 98.5 % (197/200, 95 %CI = 96.10–99.64) with LIAISON Anti-HDV. Among known HDV RNA positive samples, ARCHITECT HDV Ig detected 59/62 demonstrating 95.2 % sensitivity while LIAISON Anti-HDV sensitivity was 90.3 % (56/62). Among 101 HBV positive samples, 70 were reactive in the ARCHITECT test, 59 of which tested positive for HDV RNA for a positive predictive value (PPV) for the presence of HDV RNA was 84.3 %. For LIAISON Anti-HDV, 79 specimens were reactive and 56 contained HDV RNA: PPV for HDV RNA was 70.9 %. Among 70 HDV RNA negative samples, 39 were HBV positive. ARCHITECT HDV Ig negative predictive value (NPV) was 71.8 % and LIAISON Anti-HDV NPV was 41 % for the HBV positive group, respectively. ConclusionWhen compared to the LIASON Anti-HDV test, the ARCHITECT HDV Ig assay demonstrated enhanced sensitivity and specificity and better NPV and PPV values for HDV RNA status. The ARCHITECT HDV Ig assay represents a promising tool for universal screening of all HBsAg-positive persons.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call