Abstract

This study presents an evaluation of an advanced Doppler radar-based method for detection of vital signs, presence, and activity of a human subject in a test room with radar-signal reflecting aluminum-coated surfaces. Ten test subjects lay in four positions, and they sat in two locations in the room, both breathing normally and holding their breath. The mean ratios of the pulse rates determined from the radar signal and electrocardiography and respiration reference signals were 110% (respiration) and 99% (heartbeat), and the mean occupied and empty room radar signal variance ratios were 608 (breathing) and 20 (breath-hold). In a one-subject activity monitoring test, walking, standing and lying activities could be well separated from the radar signal. The results are promising and the proposed system seems to have potential to be used in position-independent health and activity monitoring of, for example, elderly people in care homes or intoxicated people in police custody.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.