Abstract
One aspect of the North Carolina Adult Asthma and Environment study (NCAAES) was to evaluate personal exposures to coarse particulate matter (PM 10−2.5) and their associated variability. As part of this, we examined the ability of a community-based monitor to act as a surrogate for an individual's true exposure to this size fraction in linked health effect studies. To assess personal exposures to various particulate matter (PM) size fractions, a personal PM monitor was evaluated. This monitor featured a multi-stage cascade impactor that allowed for the simultaneous collection of PM 10−2.5 and PM 2.5 size fractions. The monitor was evaluated for collocated bias and comparability with a dichotomous (dichot) sampler (device for dividing aerosol PM population into two size fractions during sampling) at an outdoor monitoring site. Results of this evaluation indicated that the prototype was capable of agreement within ±20% of that provided by the reference methodology as well as 20% daily precision for PM 10−2.5 mass measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.