Abstract

A novel X-ray system, providing dichromatic beams for dual-energy radiography, has been assembled. The source generates pairs of superimposed quasi-monochromatic beams, having energies E and 2 E, with E tuneable in the 15–20 keV range. In this paper the characteristics of the radiation field in terms of energy resolution and fluence, for three dichromatic X-ray beams are reported. A study of the spectra attenuated by a 5 cm-thick phantom of breast equivalent tissue demonstrates that the optimal energy of the dichromatic beam for dual-energy application may be set as a function of the thickness of investigated tissue. A detailed topographic study of mean energy and flux shows the spatial superposition of the first and the second diffraction order beam, that is the main requirement for the application of a single exposure dual-energy radiography. The bidimensional mapping of the irradiated beam is also reported, showing the presence of energy and intensity gradients. We estimate that the observed gradients do not affect the results of dual-energy technique application in an appreciable way.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.