Abstract

To evaluate the effectiveness of a novel crescent-shaped tibial plateau-leveling osteotomy (TPLO) saw guide (crescent guide) to assist with saw control in novice participants. Ex vivo study. Synthetic bones (n=54) and medium sized dog pelvic limbs (n=36). The 6 participants (interns and residents) without any prior experience performing a TPLO each performed 9 osteotomies on synthetic tibia models, and 6 osteotomies in cadaveric limbs of medium-sized dogs. Osteotomies made with the crescent guide were compared with those made with a standard jig and a radial saw guide with a jig. Osteotomy angulation, distance of eccentricity (DOE), and medial tibial cortical damage (synthetic bone models only) were measured from calibrated photographs. Participants rated their experiences with each technique. There was no difference in the DOE, coronal or axial osteotomy angulation between the 3 alignment devices for synthetic bone models or cadavers. Average medial cortical damage with the crescent guide (3.8 ± 7.3 mm2 ) was lower than with the radial guide (35.7 ± 27 mm2 ) and standard jig (51.2 ± 63.2 mm2 ) guides (P=<.01). Five of 6 participants preferred the crescent guide over the standard jig and radial guide. There was no difference in accuracy of osteotomy positioning but using the crescent guide resulted in lower cortical damage and more favorable participant perceptions. The crescent guide may improve control of the radial saw during TPLO in novice surgeons but does not appear to aid accurate osteotomy positioning.