Abstract

The aim of the present study was to evaluate whether equine serum amyloid A (SAA) concentrations could be measured reliably with a turbidometric immunoassay (TIA) developed for use with human serum. Intra- and inter-assay imprecision were evaluated by multiple measurements on equine serum pools. Assay inaccuracy was determined by linearity under dilution. The assay was subsequently used for measuring SAA concentrations in clinically healthy horses, horses with inflammatory diseases, horses with non-inflammatory diseases, and in horses before and after castration. In pools with low, intermediate and high SAA concentrations, the intra-assay imprecisions were 24.4%, 1.6% and 2.1%, and the inter-assay imprecisions were 33.2%, 4.6% and 6.5%. Slight signs of inaccuracy were observed, but these inaccuracies were negligible when considering the large dynamic range of the SAA response. The assay was able to detect the expected difference in SAA levels in different groups of horses. It was also able to demonstrate the expected dynamic changes in SAA after castration. In conclusion, equine SAA concentrations can be measured reliably using the TIA designed for human SAA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.